SCR SA

  • Plein écran
  • Ecran large
  • Ecran étroit
  • Increase font size
  • Default font size
  • Decrease font size
French (Fr)English (United Kingdom)

Oxygen probes SDT

E-mail Print PDF

 

SDT Probes very robust by their mechanics have been constructed to last .

Their measure element of last generation is based on a ZrO2 electrolyte, for a reliability and a very high precision

 

Sdt

         

General features :

Main tube in inconel 600 and head of probe in special refractory alloy.
Outside protective tube in thermal ceramics shock-proof.
Temperature measurement by thermocouple type S interchangeable.
Internal electode in platinium

Probe perfectly insulated
Easy maintenace

Diameter of passage 35mm
Connection on the furnace DN40 and 1 inch

Current uses : Cementation, carbonitriding, trempering…

Standards lenghts available on stock (600/700/800/900 mm)
Delivered with a test report


Technical specifications :

Signal generated by the probe :  0-1200mV
Thermocouple :  Type S or without
Lenth : 600,700,800 or 900 mm
Maximal temperature : 1150°C
Air of reference : 40 l/h
Air of purge : 150 l/h

 

 


News

NH3_cracker
- Single or double cracker for ammonia
- 3 sizes, powers 5.4, 10.8 and 21.6 kW
- Control system at 950°C secure

Ideal for destruction
residual ammonia gases and other

And/or for fabrication
process gases (H2+N2)

NH3 emissions < 0.0042 g / hour
NOx emissions = 46 mg/Nm3 sec
Complies with NF and European standards
Veritas certified

Contact us for more information

 
You are here: Products Oxygen probes Oxygen probes SDT